• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON LABORATORIES IRELAND LTD. ACRYSOF ASPHERIC UV ABSORBING SINGLEPIECE IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON LABORATORIES IRELAND LTD. ACRYSOF ASPHERIC UV ABSORBING SINGLEPIECE IOL; INTRAOCULAR LENS Back to Search Results
Model Number SA60WF
Device Problem Device Contaminated During Manufacture or Shipping (2969)
Patient Problem Insufficient Information (4580)
Event Date 09/13/2022
Event Type  malfunction  
Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.Photo provided showing implanted iol(intraocular lens) on a monitor screen.There appears to be a dark line in the center of the iol.Reported complaint cannot be confirmed from the provided photo.The manufacturer internal reference number is: (b)(4).
 
Event Description
A physician reported that during implantation of an intraocular lens (iol) fibrous tissue (looking like a scratch on the posterior optic) appeared after the iol implantation.The surgery was completed the same day.No information about lens removal or replacement.Additional information was requested and received stating the patient condition is unknown.The lens is still implanted in the patient's eye.
 
Manufacturer Narrative
The complainant indicates the use of methyl as a viscoelastic, which is not qualified for use with associated model.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF ASPHERIC UV ABSORBING SINGLEPIECE IOL
Type of Device
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15585139
MDR Text Key306608533
Report Number9612169-2022-00524
Device Sequence Number1
Product Code HQL
UDI-Device Identifier00380652332682
UDI-Public00380652332682
Combination Product (y/n)N
Reporter Country CodeEG
PMA/PMN Number
P930014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/12/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSA60WF
Device Catalogue NumberSA60WF.210
Device Lot Number21268286
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/01/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/09/2020
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
METHYL; MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D; MONARCH IOL DELIVERY SYSTEM, INJECTOR, UNSPECIFIED
-
-