Model Number 6260-9-232
Medical Device Problem Codes Degraded (1153); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Ambulation or Postural Difficulties (2544)
Date of Event 09/19/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported the patient's hip was revised due to failure of the femoral neck (rep clarified the trunnion looked ground down and there was a notch in the neck).The stem, head, and liner were revised.Rep confirmed there are no allegations against the revised liner.
 
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.
 
Event or Problem Description
It was reported the patient's hip was revised due to failure of the femoral neck (rep clarified the trunnion looked ground down and there was a notch in the neck).The stem, head, and liner were revised.Rep confirmed there are no allegations against the revised liner.
 
Additional Manufacturer Narrative
Reported event: an event regarding disassociation involving a metal head was reported.The event was confirmed through review of images provided where the stem was worn compromising the taper lock.Method & results: -device evaluation and results: visual inspection: no product was returned for evaluation however photographs were provided for review.The photographs show a recently explanted head, stem and liner.Blood is visible on all devices.Bony ingrowth is visible on the body of the stem.The trunnion is worn/corroded.Damage is visible on the rim of the liner.There is nothing of note for the head.Material analysis, functional and dimensional inspections could not be performed as the device was not returned.-clinician review: no medical records were received for review with a clinical consultant.-device history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the reported lot.Conclusions: it was reported that the patient was revised due to wear of the stem and the subsequent disassociation of the head from the stem.The event was confirmed by visual inspection of provided photographs of the devices, where the stem was worn compromising the taper lock.No further investigation for this event is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
32MM +4 LFIT V40 HEAD
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
marisol santiago
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15585386
Report Number0002249697-2022-01473
Device Sequence Number16834643
Product Code JDI
UDI-Device Identifier07613327013085
UDI-Public07613327013085
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K010757
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2013
Device Model Number6260-9-232
Device Catalogue Number6260-9-232
Device Lot Number26108102
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/19/2022
Supplement Date Received by Manufacturer12/08/2022
Initial Report FDA Received Date10/12/2022
Supplement Report FDA Received Date12/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/21/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age85 YR
Patient SexFemale
-
-