• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MATERIALISE NV TRUMATCH; TRUMATCH MIDFACE/MANDIBLE - TITANIUM 3D PRINTED PLATE FOR MANDIBLE, MEDIUM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MATERIALISE NV TRUMATCH; TRUMATCH MIDFACE/MANDIBLE - TITANIUM 3D PRINTED PLATE FOR MANDIBLE, MEDIUM Back to Search Results
Model Number SD980.020
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Osteomyelitis (4533)
Date of Event 03/25/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation ongoing.Device still in patient.
 
Event or Problem Description
Patient presented with persistent swelling and osteomyelitis/infection.Surgeons reviewed ct scans and see some bone degradation and loose screws.Surgeons suspect that the patient did not heal well from the original surgery and the infection came first and caused the loose screws (but there is no way to tell).Patient also has type ii diabetes which they think was a contributing factor.Existing plate will be removed, fibula and ramus segment slightly repositioned and then new milled plate put on.
 
Additional Manufacturer Narrative
"investigation concluded that the device met specifications and did not malfuction.It was in surgeons' opinion that the patient did not heal well leading to the infection which then caused screws loosening.An other possible contributing factor was that the patient had type ii diabetes.".
 
Event or Problem Description
"patient presented with persistent swelling and osteomyelitis/infection.Surgeons reviewed ct scans and see some bone degradation and loose screws.Surgeons suspect that the patient did not heal well from the original surgery and the infection came first and caused the loose screws (but there is no way to tell).Patient also has type ii diabetes which they think was a contributing factor.Existing plate will be removed, fibula and ramus segment slightly repositioned and then new milled plate put on.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUMATCH
Common Device Name
TRUMATCH MIDFACE/MANDIBLE - TITANIUM 3D PRINTED PLATE FOR MANDIBLE, MEDIUM
Manufacturer (Section D)
MATERIALISE NV
technologielaan 15
leuven, vlaams-brabant 3001
BE  3001
Manufacturer (Section G)
MATERIALISE NV
technologielaan 15
leuven, vlaams-brabant 3001
BE   3001
Manufacturer Contact
jenny jones
technologielaan 15
leuven, vlaams-brabant 3001
BE   3001
MDR Report Key15585853
Report Number3003998208-2022-00023
Device Sequence Number2809648
Product Code JEY
UDI-Device Identifier05420060380204
UDI-Public05420060380204
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K170272
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberSD980.020
Device Catalogue NumberSD980.020
Device Lot NumberMU22SORGUH
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/03/2022
Supplement Date Received by Manufacturer10/03/2022
Initial Report FDA Received Date10/12/2022
Supplement Report FDA Received Date11/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/16/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
Patient SexMale
-
-