The device was evaluated by olympus.The evaluation found the allegation "failed leakage test" was confirmed.The insertion bending section cover was found leaking, and the control unit was corroded.Scratches were found in multiple locations on the device.The investigation is ongoing.If additional information becomes available, this report will be supplemented accordingly.
|
The customer reported to olympus, during reprocessing, the uretero-reno videoscope did not pass the leakage test.A pinhole and damage to the outer seal were reported.There were no reports of patient harm associated with this event.During the evaluation of the device, it was found the bending section was broken with metal protruding from the bending section cover.This report is to capture the reportable malfunction of the broken bending section noted at estimation.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the reported event is unable to be determined.However, the cause of the event is likely due to the bending tube got broken by pushing with excessive force while bending section was being fixed.The event can be detected by following the instructions for use (ifu) which state: inspection of the endoscope: inspect the external surface of the entire insertion section, including the bending section and the distal end for dents, bulges, swelling, scratches, peeling of coating, holes, sagging, transformation, bends, adhesion of foreign bodies, missing parts, protruding objects, or other irregularities.The event can be prevented by following the instructions for use (ifu) which state: do not insert the insertion tube with excessive force into the ureter or calix.The bending section may be damaged.Olympus will continue to monitor field performance for this device.
|