Model Number 1504-40-208
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Pain (1994); Joint Laxity (4526); Unspecified Musculoskeletal problem (4535); Insufficient Information (4580)
Date of Event 09/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
Clinical adverse event received for patellar clunk.Event is serious and is considered moderate.Event is possibly related to procedure.Event is not related to device.Date of implant: (b)(6) 2020.Date of event: (b)(6) 2022.(right knee).Treatment: no intervention documented; awaiting scope.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint #
=
> (b)(4) this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: b5, d10 (concomitant med products), h6 health effect - clinical code if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: h6 health effect - impact code.
 
Event or Problem Description
Medical records received.On 09/18/2020 the patient had a revision right total knee arthroplasty, depuy components, including depuy patella and depuy cement, was used during this procedure.On 09/28/2022 medical records note the patient has had some increased grinding and crepitation around the right knee.The patient has some groin pain, which may be from the hip or knee.The physical examination notes ¿a little bit of thickening here¿ in the right knee.There is varus and valgus laxity in the knee.There is noticeable crepitation and patellar clunk.The patient reports feeling the knee catching which causes their quad to feel as though it wants to give.The patient is noted to be a good candidate for arthroscopic debridement, though there was no mention of one being scheduled.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary the device associated with this report was received for examination.Based on the visual analysis of the provided photographic evidence for device, it cannot be determinate a depuy's device failure or malfunction which could contribute to reported event.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation may be re-opened as necessary.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE CRS FEMORAL RT SZ 8 CEM
Common Device Name
KNEE FEMORAL
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY RAYNHAM, A DIV. OF DEPUY ORTHO 1219655
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key15587142
Report Number1818910-2022-20167
Device Sequence Number2724819
Product Code JWH
UDI-Device Identifier10603295042310
UDI-Public10603295042310
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K160700
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1504-40-208
Device Catalogue Number150440208
Device Lot NumberJ72F34
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/28/2022
Supplement Date Received by Manufacturer10/06/2022
11/20/2022
Initial Report FDA Received Date10/12/2022
Supplement Report FDA Received Date10/24/2022
11/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/06/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE CRS FEMORAL RT SZ 8 CEM; ATTUNE CRS RP INSRT SZ 8 14MM; ATTUNE CRS RP INSRT SZ 8 14MM; ATTUNE MEDIAL DOME PAT 41MM; ATTUNE POS FEM AUG SZ 8 4MM; ATTUNE POS FEM AUG SZ 8 8MM; ATUN FEM SLV M/L 40MM HALF POR; ATUN TIB SLV M/L 53MM FULL POR; DEPUY/CMW 2G; DEPUY1 GENTAMICIN CEMENT 40G
Patient Age65 YR
Patient SexMale
Patient Weight145 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-