| Model Number |
989605384711 |
| Medical Device Problem Code |
Application Program Freezes or Fails to Launch (4031)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/19/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An investigation is underway to determine the root cause of the issue.Results of the investigation will be included in a follow up report upon its completion.
|
| |
|
Event or Problem Description
|
|
The customer reported their cx50 ultrasound system displayed an error while taking measurements during a valve replacement procedure.The procedure was completed by using another ultrasound system.No patient or user was harmed as a result of the issue.The field service engineer reloaded the software to resolve the failure.
|
| |
|
Additional Manufacturer Narrative
|
|
A thorough investigation was performed; however, the system logs were not available due to corrupted software.The engineering team determined the issue was resolved with a software reinstallation.
|
| |
|
Additional Manufacturer Narrative
|
|
An addendum report is being submitted to provide the date received by manufacturer with the philips¿ become aware date.The information is in the g3 field.
|
| |
|