Model Number CXA230005
Medical Device Problem Codes Malposition of Device (2616); Migration (4003)
Health Effect - Clinical Code Aneurysm (1708)
Date of Event 09/12/2022
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2022, this patient underwent endovascular treatment of an abdominal aortic aneurysm using gore® excluder® aaa endoprosthesis and gore® excluder® conformable aaa endoprosthesis.While attempting to reposition the trunk-ipsilateral leg component, the device was pushed by the blood flow and fell out into the aneurysm.Attempt was made to push up the device but it was not successful due to the neck angulation, hence, the conformable aortic extender component (cuff) was deployed proximally.Because of the short length of the overlap, the physician attempted to place additional stent graft.In doing so, the gore® dryseal flex introducer sheath pushed the cuff device up.The cuff device migrated proximally and the left renal artery was unintentionally covered by the cuff device.No further treatment was performed because blood flow was observed on the angiography, and also the patient had hydronephrosis and poor renal artery function.The patient tolerated the procedure.The physician reported that the device was pushed up inadvertently.
 
Additional Manufacturer Narrative
Device evaluated by mfr: "other" was selected as the medical device remains implanted.Return not possible.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® CONFORMABLE AAA ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL PHOENIX 2 B/P
32470 n. north valley parkway
phoenix AZ 85085
Manufacturer Contact
samir kulovic
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key15587178
Report Number3007284313-2022-02155
Device Sequence Number16834658
Product Code MIH
UDI-Device Identifier00733132650934
UDI-Public00733132650934
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P200030
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/11/2024
Device Model NumberCXA230005
Device Catalogue NumberCXA230005
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/12/2022
Initial Report FDA Received Date10/12/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age82 YR
Patient SexFemale
-
-