| Model Number |
B1120-060 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event is estimated the device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the reported balloon rupture.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a lesion in an unspecified artery.The 12x60mm armada 35 balloon dilatation catheter (bdc) ruptured at a low pressure.Another balloon was used to complete the procedure.There was no adverse patient effect and there was no clinically significant delay in the procedure.No additional information was provided.
|
| |
|