Model Number 130760042
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/26/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the glenosphere thread broken 2x.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary the device associated with this report was returned for analysis.Visual analysis of the device revelated that the threads has stripped.The broken allegation cannot be confirmed due to no fractures or broken threads fragments are visible.The allegation is according for stripped condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DXTEND GLENOSPHERE ECC D42MM
Common Device Name
DELTA XTEND IMPLANTS : SHOULDER GLENOSPHERE
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY INT'L LTD. 8010379
st anthonys road
leeds LS11 8 DT
UK   LS11 8 DT
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key15587307
Report Number1818910-2022-20173
Device Sequence Number17934580
Product Code HSD
UDI-Device Identifier10603295027751
UDI-Public10603295027751
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K192855
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 10/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number130760042
Device Catalogue Number130760042
Device Lot Number5403771
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/26/2022
Supplement Date Received by Manufacturer11/03/2022
11/15/2022
12/15/2022
Initial Report FDA Received Date10/12/2022
Supplement Report FDA Received Date11/08/2022
11/17/2022
12/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/31/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
XTND GLENO ECC D42MM +2MM.
Patient SexUnknown
-
-