| Model Number |
70805-70 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/30/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
At this time, product has not yet been returned.An extended investigation has been performed for the reported complaint and determined that there was no indication that the product did not meet specification.Dhr (device history review) for the freestyle lite meter were reviewed and the dhr showed the freestyle lite meter passed all tests before release.Dhr (device history review) for the freestyle test strips were reviewed and the dhr showed the freestyle strips passed all tests before release.Retain testing was performed for freestyle strips, and all units performed in specification and passed.If the product is returned, the case will be re-opened, and a physical investigation will be performed.Note: the meter is designed to report readings of 20 mg/dl to 500 mg/dl.It should be noted that this meter does not give numeric value for readings less than 20 mg/dl.A "lo" display message indicates a reading less than 20 mg/dl.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
Customer reported receiving erratic readings on their adc device.Customer reported receiving readings of "lo" (below 20 mg/dl) and 92 mg/dl within 10 minutes.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported reader and test strips have been returned and investigated.Performed visual inspection on returned meter and no issues were observed.Meter powered on with button depression and test strip insertion.Control solution testing has been performed and all results were within range specification.No malfunction or product deficiency was identified.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
Customer reported receiving erratic readings on their adc device.Customer reported receiving readings of "lo" (below 20 mg/dl) and 92 mg/dl within 10 minutes.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|