• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC FUSION¿ ENT NAVIGATION SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC FUSION¿ ENT NAVIGATION SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9733560XOM
Medical Device Problem Code Device Sensing Problem (2917)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/19/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No parts have been received by the manufacturer for evaluation.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information regarding a navigation system used during a procedure.It was reported that two suctions were not able to be tracked and it appeared red on the screen.It was showing the instrument wasn't connected, however one seemed to stay connected better then the other.There was no impact on patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION¿ ENT NAVIGATION SYSTEM
Common Device Name
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key15587682
Report Number1723170-2022-01648
Device Sequence Number2724834
Product Code HAW
UDI-Device Identifier00643169518452
UDI-Public00643169518452
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K001284
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9733560XOM
Device Catalogue Number9733560XOM
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/19/2022
Initial Report FDA Received Date10/12/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/01/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-