ABBOTT VASCULAR TREK CORONARY DILATATION CATHETER Back to Search Results
Model Number 1012276-12
Medical Device Problem Codes Material Rupture (1546); Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/06/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported complaint appears to be related to operational context.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.It was reported that the balloon was not soaked prior to use.It should be noted that the coronary dilatation catheters (cdc), trek rx and mini trek rx, global, instruction for use states: submerge the balloon in sterile heparinized normal saline during balloon preparation, to activate the coating.Additionally, the device was not prepared (air aspiration) outside the anatomy.It should be noted that the coronary dilatation catheters (cdc), trek rx and mini trek rx, global, instruction for use specifies: all air must be removed from the balloon and displaced with contrast prior to inserting into the body, otherwise, complications may occur.In this case, it is unknown if the violations of the ifu caused or contributed to the reported complaint.
 
Event or Problem Description
It was reported that the procedure was to treat the proximal right coronary artery (rca) with moderate calcification and mild tortuosity.The 3.5x12 mm nc trek balloon dilatation catheter (bdc) was not soaked in saline prior to use and was prepped inside the anatomy.During the 3rd inflation to 10 atmospheres for 10 seconds, the balloon ruptured.An unspecified balloon was used to complete the procedure.There were no averse patient effects and there was no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15588260
Report Number2024168-2022-10563
Device Sequence Number11465659
Product Code LOX
UDI-Device Identifier08717648138560
UDI-Public08717648138560
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K103110
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2022
Device Model Number1012276-12
Device Catalogue Number1012276-12
Device Lot Number91120G1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/06/2022
Initial Report FDA Received Date10/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/20/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-