Restraint was received in 2 pieces - detached at the box stitch which secures the bed connecting strap.The bed connecting strap has been cut in a half as only about 33 inches left from the square tack (box x stitching); usually, the connecting strap is about 57 - 59 inches long.Lint found in the red and black hooks.Scratches and indentations found on both key-lock buckles.Fuzzy/frayed edges observed the black webbing.Labeling is present and legible.One a single 2799 restraint was returned (sold in pairs).No keys were returned.Evaluation found the square box stitch which secures the bed-connecting to the cuff are coming apart.The broken stitches are located in-between the webbing of the wrist strap and the overlapped bed-connecting strap.Close inspection of the box x stitch shows the top and bottom of the stitch remains complete/intact, and the broken or loose threads are contained within the plies of the webbing.Observing the ends of the threads through magnified glasses, there is a combination of threads that appear to have been pulled and threads that may have been cut.Customer reported patient was very agitated and aggressive which the ifu warns do not use this device with someone who has continued highly aggressive or combative behavior, self-destructive behavior, or deemed to be an immediate risk to others or to self.Instructions for use (ifu) were reviewed and found to provide adequate instructions and warnings for safe and effective use of the device.Per the ifu: before each use, check cuffs and straps for cracks, tears, and/or excessive wear or stretch, broken buckles or locks, and/or that hook-and-loop adheres securely as these may allow patient to remove cuff.Discard if device is damaged or if unable to lock.Additionally, the ifu warns: do not use this device with someone who has continued highly aggressive or combative behavior, self-destructive behavior, or deemed to be an immediate risk to others or to self.And additional or different body or limb restraints may be needed: if the patient pulls violently against the bed straps.At this time there is no evidence that a manufacturing non-conformity contributed to the reported complaint and the instructions for use were reviewed and determined to provide adequate instructions and warning for the safe and effective use of the device.Therefore, no corrective or preventative actions are necessary.All complaints are trended and reviewed by management on a monthly basis.As part of this monthly review, any excursion above the control limits for this failure mode will be assessed, documented and acted upon as warranted.Manufacturer reference file (b)(4).
|