| Model Number |
N/A |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer observed elevated advia centaur high-sensitivity troponin i (tnih) results on two samples from the same patient.The same sample was tested on an alternate method and the results were lower.The initial result was reported to the physician, who did not question the result.There were no reports that treatment was altered or postponed or adverse health consequences due to the discordant elevated advia centaur tnih results.
|
| |
|
Additional Manufacturer Narrative
|
|
An outside the united states customer observed elevated advia centaur high-sensitivity troponin i (tnih) results on two samples from the same patient that were considered discordant with the results from an alternate method.Calibration and quality control (qc) have been within specifications for tnih.The interpretation of results section of the advia centaur tnih instructions for use states: "results of this assay should always be interpreted in conjunction with patient's medical history, clinical presentation and other findings." siemens continues to investigate.Mdr 1219913-2022-00310 was filed for the same event.
|
| |
|
Additional Manufacturer Narrative
|
|
Initial mdr was filed on september 21, 2022 reporting, an outside the united states customer observed elevated advia centaur high-sensitivity troponin i (tnih) results on two samples from the same patient that were considered discordant with the results from an alternate method.Additional information - september 28, 2022.Potential causes: samples from patients, receiving preparations of mouse monoclonal antibodies for therapy or diagnosis may contain human anti-mouse antibodies (hama).Such samples may show either falsely elevated or falsely depressed values when tested with this method.Patient samples may contain heterophilic antibodies that could react in immunoassays to give falsely elevated or depressed results.This assay is designed to minimize interference from heterophilic antibodies.As per the clinical performance section in the instruction for use for tnih: "in the clinical trial, 11% of patients without an ami diagnosis had at least 1 advia centaur tnih test result above the 99th percentile (> 47.34 pg/ml (ng/l)) on 1 or more serial draws.89% of these patients were found to have 1 or more cardiac and non-cardiac conditions." there are several cardiac and non-cardiac conditions that can cause an elevation in troponin i above the 99th% without acute myocardial infarction (ami).The patient was a covid patient: there is much in the recent covid literature regarding heart problems that can show up while infected or hospitalized with covid-19.These heart problems are called "myocardial injury" meaning that cells in your heart muscle have died.Sometimes, people with myocardial injury don't have any heart disease symptoms.They may be diagnosed through blood tests or cardiac imaging.Others might have symptoms like chest pain or shortness of breath (dyspnea).Siemens continues to investigate.Mdr 1219913-2022-00310 supplemental 1 was filed for the same updates.
|
| |
|
Additional Manufacturer Narrative
|
|
Siemens filed the initial mdr 1219913-2022-00311 on september 21, 2022.Siemens filed the mdr 1219913-2022-00311 supplemental 1 on october 24, 2022.November 14, 2022 additional information: an outside the united states customer observed reproducibly elevated (>45 ng/l 99th% ifu) advia centaur xp high-sensitivity troponin i (tnih) results on two samples from the same patient compared to an alternate method and clinical evaluation.Siemens reviewed available information to evaluate for a potential product problem.Qc was in range and no other sample are affected indicating a patient/sample specific incident.Siemens suggested the customer run heterophile binding tube and/or serial dilutions to further investigate a potential interfering substance in the sample however after multiple requests this information was not provided.As per the clinical performance section in the instruction for use for tnih: "in the clinical trial, 11% of patients without an ami diagnosis had at least 1 advia centaur tnih test result above the 99th percentile (> 47.34 pg/ml (ng/l)) on 1 or more serial draws.89% of these patients were found to have 1 or more cardiac and non-cardiac conditions." there are several cardiac and non-cardiac conditions that can cause an elevation in troponin i above the 99th% without acute myocardial infarction (ami).Samples from patients receiving preparations of mouse monoclonal antibodies for therapy or diagnosis may contain human anti-mouse antibodies (hama).Such samples may show either falsely elevated or falsely depressed values when tested with this method.The limitations section of the instructions for use states: "patient samples may contain heterophilic antibodies that could react in immunoassays to give falsely elevated or depressed results.This assay is designed to minimize interference from heterophilic antibodies." no potential product issue is observed.The customer is operational.In section h6, the investigation finding, and investigation conclusion codes were updated.Mdr 1219913-2022-00310 supplemental 2 was filed for the same event.
|
| |
|