Model Number 1012278-12
Medical Device Problem Codes Entrapment of Device (1212); Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/30/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the reported separation or entrapment of device; however, the reported unexpected medical intervention appears to be related to circumstances of the procedure.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.The nc trek device referenced is filed under a separate medwatch report number.
 
Event or Problem Description
It was reported that the procedure was to treat a mid right coronary artery.For post-dilatation, a trek balloon was inserted and advanced to the lesion.Unspecified number of inflations were performed.During removal, some resistance with an unspecified source was met and the shaft separated, remaining in the anatomy.Another nc trek balloon was advanced in an attempt to retrieve the separated shaft of the first balloon.However, during attempted removal, the nc trek shaft also separated, although no resistance was met.A 3rd unspecified balloon was used and advanced past the two separated portions, inflated and pulled the separated shafts out of the patient anatomy.Nothing remains in the patient.The procedure concluded at this point.There was no adverse patient sequela and no clinically significant delay.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15591259
Report Number2024168-2022-10584
Device Sequence Number11344848
Product Code LOX
UDI-Device Identifier08717648138706
UDI-Public08717648138706
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K103110
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2025
Device Model Number1012278-12
Device Catalogue Number1012278-12
Device Lot Number20429G1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/30/2022
Initial Report FDA Received Date10/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/29/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
Patient SexFemale
-
-