• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. BIVONA CUSTOMIZED TRACHEOSTOMY TUBES; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. BIVONA CUSTOMIZED TRACHEOSTOMY TUBES; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 67PFPS45
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This mdr was generated under protocol (b)(4), as a result of warning letter cms# 617147.No lot number was provided; therefore, device history record review could not be completed.A product sample was received for evaluation.Visual and functional testing were performed.Two (2) pictures were received, according to the pictures received.It was seen that the unit has small fibers of material in the neck strap.One (1) sample was received without its original packaging inside a ziploc bag.Visual inspection was performed at 12 under normal conditions of illumination in order to detect any damage on the part.It was observed that the sample had excess of parylene.Additional it was tried to remove the excess of parylene using a glove.Parilene was easy removed from the unit.It was seen that productions personnel performs a 100% visual inspection during the assembly process in order to verify that the adhesive is free of contamination.Production personnel performs a 100% visual inspections before to cleaning the units in order to verify that parts are free of defects according to procedures.Quality inspectors takes a representative sample of the lot and performs a visual inspection in order to detect any contamination on the units and flacking or peeling of parylene was not acceptable on entire part.The most probable root causes are excess of parylene was produced during the superslick process since the racks was not clean correctly.The excess of parylene was produced during the superslick process since the quartz wasn?t clean correctly and it was dirty.The following actions were taken per previous complaints, as correctly actions, 6 quartz where purchased to have a backup and to make sure they had the time to clean them well.Per previous complaint, procedure was updated in order to include as part of the cleaning of the superslick, the cleaning of the rack that support the delring fixtures to prevent particulates on pieces.Per previous complaint, as mitigation a procedure was generated to ensure that racks has to be clean of parylene fibers to avoid contaminations on the units.As preventive action, production personnel was notified by area trainer on as awareness of the defect reported by the customer.
 
Event or Problem Description
It was reported that the trach never made it through first process.The technician noticed during pre-steam that the silicone on the neck flange began flaking and peeling off.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA CUSTOMIZED TRACHEOSTOMY TUBES
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15591302
Report Number3012307300-2022-23582
Device Sequence Number19350827
Product Code JOH
UDI-Device Identifier10351688518552
UDI-Public10351688518552
Combination Product (Y/N)N
PMA/510(K) Number
K923878
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number67PFPS45
Device Catalogue Number67PFPS45
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2018
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/21/2022
Initial Report FDA Received Date10/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-