MAUDE Adverse Event Report: SMITH & NEPHEW, INC. R3 LNR MPACTOR HD II 28MM; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. R3 LNR MPACTOR HD II 28MM; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL Back to Search Results
Model Number 71366428
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/20/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during inspection, two (2) r3 lnr mpactor hd ii 28mm were found to be broken.As this was noticed upon field inspection, there was no patient involvement.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
H10: additional information received by the manufacturer identified that this event should be re-evaluated for mdr reporting.The new information received confirmed that there is nothing broken on or off the r3 lnr mpactor hd ii 28mm.It did confirm multiple scratches, burrs and deep gouges in the device.The reassessment determined that the issue does not meet the threshold for reporting and therefore, it is a non-reportable event.If further details are provided, our files will be updated accordingly, and a further report submitted outlining both the event details and our investigations performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 LNR MPACTOR HD II 28MM
Common Device Name
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15591369
Report Number1020279-2022-04400
Device Sequence Number2277695
Product Code JDG
UDI-Device Identifier03596010656056
UDI-Public03596010656056
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K123598
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71366428
Device Catalogue Number71366428
Device Lot Number10KSY0035
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/20/2022
Supplement Date Received by Manufacturer10/24/2022
11/01/2022
Initial Report FDA Received Date10/12/2022
Supplement Report FDA Received Date10/31/2022
11/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/31/2010
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-