| Model Number |
115-401 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Codes |
Seroma (2069); Swollen Lymph Nodes/Glands (4432)
|
| Date of Event |
09/09/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).A review of the device history record has been completed.No deviations or non-conformance noted.The events of seroma-late and lymphadenopathy are physiological complications and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: seroma-late, lymphadenopathy, and rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side "late seroma", rupture, "pain", "implant looks higher", "swelling for five days", "enlarged, tightness, hard", and "several small axillary lymph nodes".Device remains implanted.
|
| |
|
Event or Problem Description
|
|
The device has been explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side "late seroma", rupture, "pain", "implant looks higher", "swelling for five days", "enlarged, tightness, hard", and "several small axillary lymph nodes".Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: seroma-late: unable to observe since it is a medical event and is not related to the device.Rupture: observed broken in posterior side assessed as unidentified(tear) opening.( shell thickness within specification).Pain: unable to observe since it is a medical event and is not related to the device.Visibility/palpability: unable to observe since it is a medical event and is not related to the device.Edema: unable to observe since it is a medical event and is not related to the device.Malposition: unable to observe since it is a medical event and is not related to the device.Lymphadenopathy: unable to observe since it is a medical event and is not related to the device.Additional observations: crease fold observed on the device.
|
| |
|