ABBOTT MEDICAL 8CH INFINITY DBS LEAD KIT, 30CM, 0.5, B Back to Search Results
Model Number 6170
Medical Device Problem Codes Fracture (1260); High impedance (1291)
Health Effect - Clinical Code Inadequate Pain Relief (2388)
Date of Event 09/21/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Event date is estimated.
 
Event or Problem Description
It was reported that the patient was experiencing high impedances, and then it was later discovered that the lead was fractured.Surgical intervention was undertaken and the lead was explanted and replaced.
 
Additional Manufacturer Narrative
Conclusion: the report event of impedance issue was not confirmed.As received, visual inspection and resistance testing showed the returned lead passed the functional test.The cause of the reported event remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
8CH INFINITY DBS LEAD KIT, 30CM, 0.5, B
Common Device Name
DBS LEAD
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key15591518
Report Number1627487-2022-05605
Device Sequence Number2327730
Product Code MHY
UDI-Device Identifier05415067030306
UDI-Public05415067030306
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P140009
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/30/2021
Device Model Number6170
Device Catalogue Number6170
Device Lot Number7252650
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/21/2022
Supplement Date Received by Manufacturer12/16/2022
Initial Report FDA Received Date10/12/2022
Supplement Report FDA Received Date12/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/31/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age69 YR
Patient SexFemale
-
-