| Model Number |
6170 |
| Medical Device Problem Codes |
Fracture (1260); High impedance (1291)
|
| Health Effect - Clinical Code |
Inadequate Pain Relief (2388)
|
| Date of Event |
09/21/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Event date is estimated.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was experiencing high impedances, and then it was later discovered that the lead was fractured.Surgical intervention was undertaken and the lead was explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: the report event of impedance issue was not confirmed.As received, visual inspection and resistance testing showed the returned lead passed the functional test.The cause of the reported event remains unknown.
|
| |
|