• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. CLEO; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. CLEO; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 21-7231-24
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/06/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the sett up of the catheter, the adhesive diaphragms remained stuck to the patient's belly.The needle remained in the patient's skin.No medical or surgical intervention was reported.
 
Additional Manufacturer Narrative
This mdr was generated under protocol b10009704 as a result of warning letter cms# 617147.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample was received for evaluation.Visual and functional testing were performed.Visual inspection found the sample was received in used condition without its original packaging.It was observed that the sample has the retractor active and without the needle in the retractor.During a review of the production run, the needle assembly operation was audited and it was performed by the production personnel according to the manufacturing procedures.Training records were reviewed; operators were trained on the operation been performed.Maintenance records from the machine used for the operation were reviewed; no anomalies were detected.During functional testing, no discrepancies were found.The sample was also checked to confirm that it had adhesive, comparing it with a piece of raw material without adhesive.When comparing both samples, it was observed that a piece with adhesive forms a conical part in the upper part of the component.The retractor assemblies are one hundred percent visually inspected by the production personnel to assure that the needle is properly threaded.Production personnel verifies one hundred percent that parts not be damaged (including scratches) or distorted/warped.The root cause of the reported issue was found to be the adhesive was not placed correctly.An alert was generated to ensure that operators verify that the amount of adhesive drop is correct and that it is placed as indicated in the procedure.Production personnel was also retraining to the on procedure to reinforce the needle bonding operation.(510k) is unknown.No product information has been provided to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEO
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15591688
Report Number3012307300-2022-23612
Device Sequence Number16710017
Product Code FPA
UDI-Device Identifier30610586028407
UDI-Public30610586028407
Combination Product (Y/N)N
Initial Reporter CountryFR
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 10/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/04/2022
Device Model Number21-7231-24
Device Catalogue Number21-7231-24
Device Lot Number77X187
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/11/2018
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/30/2022
Initial Report FDA Received Date10/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-