ALLERGAN (COSTA RICA) INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number TRLP-340
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 07/20/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Healthcare provider reported a right side exchange from textured to smooth breast implants due to the patient¿s concern with the product.Later, healthcare professional reported, "capsular contracture worse on the right" (baker grade unspecified,) "double capsule partially developed," "cystic lesions," and "brown greenish fluid coming out." the device has been explanted and replaced.
 
Additional Manufacturer Narrative
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade unknown.
 
Event or Problem Description
Healthcare provider reported a right side exchange from textured to smooth breast implants due to the patient¿s concern with the product.Later, healthcare professional reported, "capsular contracture worse on the right" (baker grade unspecified,) "double capsule partially developed," "cystic lesions," and "brown greenish fluid coming out." the device has been explanted and replaced.
 
Additional Manufacturer Narrative
Device evaluation: the device related to the reported event of anxiety-product/procedure, capsular contracture, double capsule and cyst-ndr was received on (b)(6)2022, with lot number 3044664.Based on the device analysis grid, the assessments of the complaint are: anxiety-product/procedure: unable to observe as it is a medical event that is not related to the device.Capsular contracture: unable to observe as it is a medical event that is not related to the device.Double capsule: unable to observe as it is a medical event that is not related to the device.Cyst-ndr: not applicable as the event is non device related.Additional observations: wear abrasion is observed.Crease flat is observed.No further actions are required as the device was implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
terry ingram
12331-a riata trace parkway
building 3
austin, TX 78727
8479366324
MDR Report Key15592016
Report Number9617229-2022-19246
Device Sequence Number11316601
Product Code FTR
UDI-Device Identifier10888628007925
UDI-Public10888628007925
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/27/2022
Device Model NumberTRLP-340
Device Catalogue NumberTRLP-340
Device Lot Number3044664
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/19/2022
Supplement Date Received by Manufacturer10/21/2022
Initial Report FDA Received Date10/12/2022
Supplement Report FDA Received Date11/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/27/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient SexFemale
Patient Weight65 KG
-
-