• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC THERMOCOOL® SF NAV BI-DIRECTIONAL CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC THERMOCOOL® SF NAV BI-DIRECTIONAL CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Model Number BNI35DFH
Medical Device Problem Codes Delivered as Unsterile Product (1421); Tear, Rip or Hole in Device Packaging (2385); Packaging Problem (3007)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 4.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster inc., or its employees that the report constitutes an admission that the product, biosense webster inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
 
Event or Problem Description
It was reported that a patient underwent a procedure with a thermocool® sf nav bi-directional catheter wherein product sterility was compromised.It was reported the packaging of the catheter was ripped, so the catheter wasn¿t sterile anymore.They had to open a new one.There were no patient consequences.Additional information received indicated the cardboard packaging was okay but the inner packaging (white back and clear front) was torn open.There was no damage to the pouch seal per se but it looked like someone opened it before.The catheter was not used on the patient because it wasn¿t sterile anymore.It looked fine though, so no obvious damage.
 
Additional Manufacturer Narrative
On 18-oct-2022, the contact information of the head of heart centre was provided and appropriate e1.Initial reporter fields have been populated.It was reported, the actual user is unknown.E1.Initial reporter phone: (b)(4) on 19-oct-2022, the bwi product analysis lab received the complaint device for evaluation.The product analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.Manufacturer's ref.# (b)(r4) on 13-nov-2022, it was determined the incorrect h6.Medical device problem codes of ¿packaging problem (a0205)¿ and ¿delivered as unsterile product (a020701)¿ were reported.The event is better represented with code ¿tear, rip or hole in device packaging (a020504)¿ as such has been updated to the appropriate field.
 
Additional Manufacturer Narrative
It was reported that a patient underwent a procedure with a thermocool® sf nav bi-directional catheter wherein product sterility was compromised.It was reported the packaging of the catheter was ripped, so the catheter wasn¿t sterile anymore.They had to open a new one.There were no patient consequences.Additional information received indicated the cardboard packaging was okay but the inner packaging (white back and clear front) was torn open.There was no damage to the pouch seal per se but it looked like someone opened it before.The catheter was not used on the patient because it wasn¿t sterile anymore.It looked fine though, so no obvious damage.Device evaluation details: the thermocool® sf nav bi-directional catheter was returned to biosense webster inc (bwi) for evaluation and the evaluation has been completed.A visual inspection of the returned device was performed following bwi procedures.Visual analysis revealed no damages or anomalies on the device.Since the original package was not returned, no analysis could be performed.A manufacturing record evaluation was performed and no internal action was found during the review.As part of biosense webster¿s quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# pc-001199793.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL® SF NAV BI-DIRECTIONAL CATHETER
Common Device Name
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
michelleann garcia
31 technology dr
irvine, CA 92618
646591-7981
MDR Report Key15592763
Report Number2029046-2022-02505
Device Sequence Number2315328
Product Code LPB
UDI-Device Identifier10846835003215
UDI-Public10846835003215
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P030031/S025
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/14/2024
Device Model NumberBNI35DFH
Device Catalogue NumberBNI35DFH
Device Lot Number30563955L
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2022
Supplement Date Received by Manufacturer10/18/2022
11/18/2022
Initial Report FDA Received Date10/12/2022
Supplement Report FDA Received Date11/15/2022
12/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
THERMOCOOL SF CARTOXP,D-F,TC.
Patient SexUnknown
-
-