STRYKER GMBH PROXIMAL PHALANGEAL BROACH, SIZE 1 SR PIP; SEMI-CONSTRAINED METAL/POLYMER FINGER JOINT PROSTHESIS Back to Search Results
Model Number 17-1325
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 07/14/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
As reported: "the size 1 proximal broach tip broke off in the patients phalange while the surgeon was trying to remove the broach.Sequential broaching (up to size 2) was completed.The tip remained in the patient as removal would have been significantly more invasive." there was a 5 minute delay to the procedure.
 
Additional Manufacturer Narrative
The reported event could be confirmed, since the device was returned and is indeed broken.The device inspection could confirm the catalog and lot numbers reported were correct, and that tip of the broach is indeed broken.The device shows various scratches, which could be attributed to multiple years of use and reprocessing.The tip of the device shows a brittle fracture, showing that the instrument broke in a sudden manner, following the application of an important force.Based on investigation, the root cause was attributed to a user related issue.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If more information is provided, the case will be reassessed.
 
Event or Problem Description
As reported: "the size 1 proximal broach tip broke off in the patients phalange while the surgeon was trying to remove the broach.Sequential broaching (up to size 2) was completed.The tip remained in the patient as removal would have been significantly more invasive." there was a 5 minute delay to the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROXIMAL PHALANGEAL BROACH, SIZE 1 SR PIP
Common Device Name
SEMI-CONSTRAINED METAL/POLYMER FINGER JOINT PROSTHESIS
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ   2545
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15593974
Report Number0008031020-2022-00542
Device Sequence Number13987316
Product Code MPK
UDI-Device Identifier00886385019840
UDI-Public00886385019840
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number17-1325
Device Catalogue Number171325
Device Lot Number038801
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/29/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/19/2022
Supplement Date Received by Manufacturer02/22/2023
Initial Report FDA Received Date10/13/2022
Supplement Report FDA Received Date03/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/10/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexFemale
-
-