• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE RP TIB BASE SZ 6 CEM; ATTUNE IMPLANT : KNEE TIBIAL TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE RP TIB BASE SZ 6 CEM; ATTUNE IMPLANT : KNEE TIBIAL TRAY Back to Search Results
Model Number 1506-10-006
Medical Device Problem Codes Loss of or Failure to Bond (1068); Migration (4003)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/27/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Patient was revised due to aseptic tibial and femoral loosening.Tibial tray had also fallen into varus.Femoral component loosened at the cement/bone interface.Tibial component loosened at the cement/implant interface.Tibial tray¿s underside was completely clean of cement when removed.Cement manufacturer was unknown.There was no surgical delay.Doi: (b)(6) 2014; dor: (b)(6) 2022; affected side: right knee.
 
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE RP TIB BASE SZ 6 CEM
Common Device Name
ATTUNE IMPLANT : KNEE TIBIAL TRAY
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key15594534
Report Number1818910-2022-20320
Device Sequence Number2136701
Product Code NJL
UDI-Device Identifier10603295042600
UDI-Public10603295042600
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1506-10-006
Device Catalogue Number150610006
Device Lot Number3498008
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/27/2022
Supplement Date Received by Manufacturer10/31/2022
Initial Report FDA Received Date10/13/2022
Supplement Report FDA Received Date11/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/31/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE PS FEM RT SZ 7 CEM; ATTUNE PS RP INSRT SZ 7 7MM; ATTUNE RP TIB BASE SZ 6 CEM; BONE CEMENT (MFG. AND PRODUCT NAME UNKNOWN)
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
Patient SexMale
-
-