| Model Number |
TN FC500 FLOW CYTOMETER 100-240V 50/60HZ |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/05/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient information is not applicable.There was no impact to patients as a result of this event.The field service engineer (fse) was on site and confirmed what the customer reported.To resolve the issue the fse replaced the tarpon amp boards at the forward scatter (fs), and fl5 positions.Bec is filing an mdr for this event based on the fda classification of the (b)(6) 2018 urgent medical device recall as a class i recall on (b)(6)2018 (recall number z-0471-2019 for fc 500; recall number z-0472-2019 for epics xl/xlmcl).Bec internal identifier - (b)(4).
|
| |
|
Event or Problem Description
|
|
The customer reported signal drift at the forward scatter (fs), and fl5 positions on their fc 500 flow cytometer while running quality control.There was no report of death, injury, or change to patient treatment as a result of this event.
|
| |
|