• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECKMAN COULTER FC 500 FLOW CYTOMETER; COUNTER, DIFFERENTIAL CELL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECKMAN COULTER FC 500 FLOW CYTOMETER; COUNTER, DIFFERENTIAL CELL Back to Search Results
Model Number TN FC500 FLOW CYTOMETER 100-240V 50/60HZ
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/05/2022
Event Type  malfunction  
Manufacturer Narrative
Patient information is not applicable.There was no impact to patients as a result of this event.The field service engineer (fse) was on site and confirmed what the customer reported.To resolve the issue the fse replaced the tarpon amp boards at the forward scatter (fs), and fl5 positions.Bec is filing an mdr for this event based on the fda classification of the (b)(6) 2018 urgent medical device recall as a class i recall on (b)(6)2018 (recall number z-0471-2019 for fc 500; recall number z-0472-2019 for epics xl/xlmcl).Bec internal identifier - (b)(4).
 
Event Description
The customer reported signal drift at the forward scatter (fs), and fl5 positions on their fc 500 flow cytometer while running quality control.There was no report of death, injury, or change to patient treatment as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FC 500 FLOW CYTOMETER
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 s. kraemer blvd.
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th avenue
miami FL 33196 2031
Manufacturer Contact
angela vettel
250 s. kraemer blvd.
brea, CA 92821
3053802874
MDR Report Key15594938
MDR Text Key307079048
Report Number1061932-2022-00064
Device Sequence Number1
Product Code GKZ
UDI-Device Identifier15099590572839
UDI-Public(01)15099590572839(11)NO-DATA
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K071681
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Remedial Action Notification
Type of Report Initial
Report Date 10/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTN FC500 FLOW CYTOMETER 100-240V 50/60HZ
Device Catalogue Number626553
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/05/2022
Initial Date FDA Received10/13/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction Number012-0108/2018-001C
Patient Sequence Number1
-
-