This is filed to report a leak.It was reported that on 20sep2022, during preparation of a g4 steerable guide catheter (sgc), there was a loss of column observed.This occurred while inverting the sgc, with the finger removed from the tip, hand fully emerged, and the stopcock was closed.The chamber below the hemostasis valve was leaking.The sgc was replaced.There was no patient involvement.No additional information was provided.
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to the reported event.Additionally, a review of the complaint history identified no other complaints from the lot.All available information was investigated and without the device to analyze, a cause for the reported leak/splash associated with loss fluid column during device preparation could not be determined.There is no indication of a product issue with respect to manufacture, design or labeling.
|