A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: visual analysis of the returned device identified: broken plug strap, crease fold, part of the shell is missing.Leak test was performed and found broken on anterior.A microscopic analysis was performed which identify sharp edge broken assessed as unidentified tear broken.A dimension measurement in the shell was performed which identify the thickness within specification.The fill test inspection was performed, the result is no blockage.Based on the device analysis the final assessment is: a sharp broken on anterior assessed as unidentified (tear) broken.Further information from the reporter regarding event, product, or patient details has been/will be requested.No additional information is available at this time.Reason for reoperation: deflation.
|