Model Number 72203290
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/30/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that during a labrum repair surgery the osteoraptor anchor broke while implantation.The procedure was finished with a smith and nephew back up device.It is unknown if a surgical delay occurred to this event.No further complications were reported.
 
Additional Manufacturer Narrative
H3, h6: the reported device was received for evaluation.There was a relationship found between the device and the reported event.A complaint history review found similar reported events.Insufficient product identification information was provided, and thus, a manufacturing record review could not be conducted.Based on the information available, there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the polymer, found that the storage requirements, material specifications, and applicable tests were appropriately specified.A material certificate of analysis was required for the raw material.A visual inspection of the returned device found that it is not in its original packaging.The implantation device and a fractured anchor, with the suture material through the suture window were returned.The distal end of the implantation device is deformed and the anchor is fractured just below the suture window.All of the returned items have debris on them.Based on the condition of the product material found during visual inspection, additional material testing is not required.This case reports the breakage of the osteoraptor anchor and it is unknown if the broken pieces were retrieved from the patient.An undated photo of the inserter was provided for review, the anchor was not in the photo.Based on the information provided no clinical factors were found which would have contributed to the reported breakage.The osteoraptor anchor is implantable, biocompatibility is not an issue.If retained micro-motion or movement cannot be predicted.There is potential risk for tissue inflammation and/or pain.No further clinical/medical assessment is warranted at this time.The complaint was confirmed, and the root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include excessive force on the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event inconsistent with normal use.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
 
Additional Manufacturer Narrative
H10: additional information.Updated: b2, b5 and h6 (health effect - impact code).
 
Event or Problem Description
It was reported that, during a labrum repair surgery, the osteoraptor anchor broke while implantation.All the broken pieces were successfully retrieved with suture.The procedure was finished with a smith and nephew back up device, used in the originally drilled bone hole.No delay or further complications were reported.The health status of the patient is good.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSTEORAPTOR CRVD 2.3 SA UB COBRD BLUE
Common Device Name
STAPLE, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15595954
Report Number1219602-2022-01405
Device Sequence Number2314894
Product Code JDR
UDI-Device Identifier00885554022254
UDI-Public00885554022254
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K151105
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72203290
Device Catalogue Number72203290
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/30/2022
Supplement Date Received by Manufacturer10/07/2022
02/03/2023
Initial Report FDA Received Date10/13/2022
Supplement Report FDA Received Date10/19/2022
02/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age55 YR
Patient SexFemale
Patient Weight60 KG
-
-