• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® INTRA-ABDOMINAL PRESSURE MONITORING DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® INTRA-ABDOMINAL PRESSURE MONITORING DEVICE Back to Search Results
Model Number IAP001
Medical Device Problem Code Incomplete or Missing Packaging (2312)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/29/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the intra abdominal pressure measurement kit came in packaged a bit differently and it had caused some confusion.Stated that the syringe came unattached which caused a confusion.
 
Event or Problem Description
It was reported that the intra abdominal pressure measurement kit came in packaged a bit differently and it had caused some confusion.Stated that the syringe came unattached which caused a confusion.
 
Additional Manufacturer Narrative
The reported event was confirmed cause unknown.1sample were confirmed to exhibit the reported failure.The device had not met specifications.The product was not used for patient treatment.The product caused the reported failure.Visual evaluation of the returned photo sample noted one unopened (with original packaging), intra-abdominal pressure measurement kit.The device history record review could not be performed without a lot number.The investigation is concluded, and no additional action is required at this time.The instructions for use were found adequate and state the following: "indications for use: the bard® intra-abdominal pressure (iap) monitoring device is intended for the monitoring of intra-abdominal pressure via a foley urinary catheter.The measured pressures can be used as an aid in the diagnosis of intra-abdominal hypertension (iah) and the associated clinical syndrome of abdominal compartment syndrome (acs).Caution as an interpretive tool, the bard® intra-abdominal pressure monitoring device should be used along with other clinical indicators to aid the physician in the diagnosis of intra-abdominal hypertension (iah) and the associated clinical syndrome of abdominal compartment syndrome (acs).Device description: the bard® intra-abdominal pressure monitoring device is composed of a tubing set used for infusing fluid into the urinary bladder through the foley catheter sampling port.It utilizes a clamping device to occlude the urinary drainage tubing to form a fluid column through which pressure is measured.Important: the bard® intra-abdominal pressure monitoring device does not include a saline bag, pressure transducer or monitoring system.It adapts to the saline bag, pressure transducer and monitoring system used in your icu.The bard® intra-abdominal pressure monitoring device must be replaced at the time the urinary catheter and/or urinary drainage tubing is replaced.The bard® intra-abdominal pressure monitoring device is for use with bard® foley trays with ez-lok® sampling port.Contraindications not for use on pediatric patients not for use on patients with compromised bladder function warnings use only saline for pressure measurement.Ensure tubing fluid path is primed so there are no air bubbles.Ensure the iap sampling port is seated tightly to the catheter drainage tubing sampling port prior to use.Precautions single use only federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Not made with natural rubber latex sterile unless package is opened or damaged, except for any individually packaged components within the tray which are not labeled as sterile.These components are not terminally sterilized.This is a single use device.Do not resterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Sterile contents: 30 cc syringe infusion and pressure transducer tubing saline bag spike & cap zeroing stopcock dead-end transducer cap proprietary drain tube clamp valve port check valves for controlling and directing flow statlock® foley device kit." the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® INTRA-ABDOMINAL PRESSURE MONITORING DEVICE
Common Device Name
INTRA-ABDOMINAL PRESSURE MONITORING DEVICE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key15596054
Report Number1018233-2022-07808
Device Sequence Number16694448
Product Code PHU
UDI-Device Identifier00801741121098
UDI-Public(01)00801741121098
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K070201
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIAP001
Device Catalogue NumberIAP001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/29/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/29/2022
Supplement Date Received by Manufacturer02/27/2023
Initial Report FDA Received Date10/13/2022
Supplement Report FDA Received Date02/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-