Model Number FT18DN40NSA053N
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/14/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This mdr was generated under protocol (b)(4), as a result of warning letter cms# 617147.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample was received for evaluation.Visual and functional testing were performed.The returned device was visually inspected and the device was confirmed broken near the neck flange/airway junction.The reported problem was confirmed.It appears the airway line split away from the shaft perhaps from excessive pulling or it came in contact with something sharp that would cause a cut/tear.The root cause cannot be confirmed since what exactly caused the splitting of the shaft during use cannot be determined.Actions were taken to mitigate the reported issue: monitor this type of complaints or trends.Initial reporter also sent report to fda is unknown.No information has been provided to date.
 
Event or Problem Description
It was reported that the upon opening the device, a hole in the cuff was observed.No patient injury was observed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA TRACHEOSTOMY TUBE, NEONATAL/PEDIATRIC
Common Device Name
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15597211
Report Number3012307300-2022-23833
Device Sequence Number2072224
Product Code JOH
UDI-Device Identifier15019517147193
UDI-Public15019517147193
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K914088
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFT18DN40NSA053N
Device Catalogue NumberFT18DN40NSA053N
Device Lot NumberDS015643
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/09/2018
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/06/2022
Initial Report FDA Received Date10/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age5 MO
Patient SexMale
Patient Weight5 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-