This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on october 13, 2022.Upon further investigation of the reported event, the following information is new and/or changed: g6 (indication that this is a follow-up report).H2 (follow-up due to additional information and device evaluation).H6 (identification of evaluation codes 10, 3331, 3224, 3259, 19).Type of investigation #1: 10 - testing of actual/suspected device.Type of investigation #2: 3331 - analysis of production records.Investigation findings #1: 3224 - optical problem identified.Investigation findings #2: 3259 - improper physical structure.Investigation conclusions: 19 - cause traced to user.The returned sample was visually inspected for damage and looked through the endoscope directly to read some print matter.The internal components was then inspected using a monocular, and was discovered that there was a crack on one of the internal lenses which resulted in a blurry visual field.Improper handling caused the internal lens to break.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
|