| Model Number |
HT18FS50NGF113N |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/25/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during the cuff integrity check of the device it was found that the cuff was leaking, upon closer investigation the cuff appeared to be torn away.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).There were no non-conforming reports (ncrs) initiated for the lot or anomalies identified in the device history record (dhr) review.One device received at the investigation site for inspection.Visual inspection was performed with the unaided eye under normal plant lighting and the reported issue of damaged cuff was confirmed.The investigation did not reveal any manufacturing defects with the device.No corrective actions are planned at this time.The root cause was unable to be determined.
|
| |
|