Model Number SLActive Titanium TL
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Code Fibrosis (3167)
Date of Event 09/02/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The batch number could be verified.Our manufacturing q-system assures, that production and process controls are in place to ensure that batches confirm to the applicable specifications before they are distributed.  the loss of an endosseous dental implant after successful osseointegration and restoration is a known inherent risk of the procedure.Implants may have to be removed in case one or more of the implant success criteria are not met.Implant success criteria according to buser et al.(1991) are: 1.Absence of persistent subjective complaints such as pain, foreign body sensation and /or dysesthesia.2.Absence of a recurrent peri-implant infection with suppuration.3.Absence of implant mobility.4.Absence of a continuous radiolucency around the implant.The manufacturer's trend analysis confirms that the reported failure rate associated with its dental implants is below the expected failure rate for this treatment as published in the scientific literature.
 
Event or Problem Description
The clinician reports the implant was inserted (b)(6) 2010 in fdi 26.On (b)(6) 2022, loss of osseointegration was verified.The device was forwarded to the manufacturer.At the event the patient experienced: mobility.No further patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S-IMPLANT Ø 4.8MM WN, SLACTIVE 6MM
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
INSTITUT STRAUMANN AG
peter merian-weg 12
basel 04002
SZ  04002
Manufacturer (Section G)
INSTITUT STRAUMANN AG
peter merian-weg 12
basel 04002
SZ   04002
Manufacturer Contact
jennifer jackson
60 minuteman road
andover, MA 01810
9787472509
MDR Report Key15598438
Report Number0009613348-2022-93735
Device Sequence Number2200683
Product Code DZE
UDI-Device Identifier07630031716323
UDI-Public07630031716323
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K033984
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 10/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2014
Device Model NumberSLActive Titanium TL
Device Catalogue Number033.630S
Device Lot NumberR3727
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/19/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date10/13/2022
Event Location Other
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date10/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/28/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
Patient SexFemale
-
-