• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. BIVONA TUBES CUSTOM; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. BIVONA TUBES CUSTOM; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number FT18DN65NSC129S
Medical Device Problem Codes Break (1069); Material Too Rigid or Stiff (1544)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/16/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
There were no non-conforming reports (ncrs) initiated for the lot or anomalies identified in the device history record (dhr) review.No product sample was received; therefore, visual and functional testing could not be performed.The reported issue could not be confirmed as no product sample was received for evaluation.If the product is returned, the manufacturer will reopen this complaint for further investigation.This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Event or Problem Description
It was reported that the device's swivel connector was too tight, was extremely difficult to remove, the ring broke and there was a kind of bulge in the cannula at the point where the cuff starts.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA TUBES CUSTOM
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15598581
Report Number3012307300-2022-23865
Device Sequence Number19402727
Product Code JOH
UDI-Device Identifier15019517148916
UDI-Public15019517148916
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K923878
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFT18DN65NSC129S
Device Catalogue NumberFT18DN65NSC129S
Device Lot NumberDS017653
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/07/2022
Initial Report FDA Received Date10/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-