• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM ONBOARD BATTERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM ONBOARD BATTERY Back to Search Results
Catalog Number 295025-001
Medical Device Problem Code Battery Problem (2885)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/20/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The freedom onboard battery will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.
 
Event or Problem Description
The freedom onboard battery was not supporting a patient at the time of the event.The customer, a syncardia certified hospital, reported that during a training event the freedom onboard battery did not have a charge.The hospital attempted to charge the battery without success.
 
Additional Manufacturer Narrative
Device history record (dhr) review confirmed that freedom onboard battery s/n (b)(6) was serviced and passed all functional testing prior to being released to finished goods.Alarm history and data file review found no anomalies or permanent faults observed.Visual inspection found separation in the battery housing.Batteries passed all areas of functional testing.Battery was rejected ue to cosmetic damage of the housing, although damage did not affect functionality.Complaint could not be replicated.Failure investigation for this complaint could not confirm the reported issue.The customer complaint was not replicated during testing; the root cause of the reported inability to charge the batteries was unable to be determined.Failure investigation identified no test failures or damage that could have contributed to the complaint.The damage found was cosmetic only.No evidence of a device malfunction was found.Batteries were not in patient use at time of complaint.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its evaluation and is closing this file.If new or additional information is received in the future, syncardia will file a follow-up mdr.
 
Event or Problem Description
The freedom onboard battery was not supporting a patient at the time of the event.The customer, a syncardia certified hospital, reported that during a training event the freedom onboard battery did not have a charge.The hospital attempted to charge the battery without success.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM ONBOARD BATTERY
Common Device Name
BATTERY
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e silverlake road
tucson AZ 85713
Manufacturer (Section G)
SYNCARDIA SYSTEMS, LLC
1992 e silverlake road
tucson AZ 85713
Manufacturer Contact
aaron meier
1992 e silverlake road
tucson, AZ 85713
5205451234
MDR Report Key15598825
Report Number3003761017-2022-00107
Device Sequence Number11332346
Product Code LOZ
UDI-Device Identifier00858000003084
UDI-Public(01)00858000003084
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P030011
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 06/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number295025-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/21/2022
Supplement Date Received by Manufacturer09/21/2022
Initial Report FDA Received Date10/13/2022
Supplement Report FDA Received Date06/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/12/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-