PENUMBRA, INC. PENUMBRA SYSTEM 3MAX REPERFUSION CATHETER; NRY Back to Search Results
Catalog Number 3MAXC
Medical Device Problem Codes Break (1069); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/17/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Evaluation of the returned 3maxc confirmed the reported complaint.Evaluation revealed that the device was stretched and fractured on the distal shaft.If the device is retracted against resistance, damage such as this may occur.Based on the reported complaint, the root cause of resistance during the procedure could not be determined.Further evaluation revealed kinks on the distal fractured segment.This damage was likely incidental and may have occurred during snaring of the device during removal.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
The patient was undergoing a thrombectomy procedure in the middle cerebral artery (mca) using a penumbra system 3max reperfusion catheter (3maxc), a neuron max 6f 088 long sheath (neuron max), and a guidewire.During the procedure, the physician inserted the neuron max into the patient and then advanced the 3maxc into the target vessel.The physician then completed the first pass using the 3maxc.While removing the 3maxc, the physician experienced resistance and consequently, the distal end of the 3maxc broke off.Therefore, the physician used a snare device to remove the broken part of the 3maxc from the patient.The procedure was completed using a new 3maxc and the same neuron max.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM 3MAX REPERFUSION CATHETER
Common Device Name
NRY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key15598856
Report Number3005168196-2022-00472
Device Sequence Number11341751
Product Code NRY
UDI-Device Identifier00814548019956
UDI-Public814548019956
Combination Product (Y/N)N
Initial Reporter CountryMC
PMA/510(K) Number
K160449
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/13/2022
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/15/2024
Device Catalogue Number3MAXC
Device Lot NumberF110779
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/30/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/20/2022
Initial Report FDA Received Date10/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
Date Report Sent to FDA01/01/2005
-
-