• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. BIVONA TRACHEOSTOMY TUBE, ADULT; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. BIVONA TRACHEOSTOMY TUBE, ADULT; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 67HA90
Medical Device Problem Code Device Markings/Labelling Problem (2911)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/08/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that upon opening the product, the nurse read the product instruction for use (ifu) and compared the english text to the finnish text and noticed that the translation was not correct, observed with four lot numbers.No patient injury was reported.
 
Additional Manufacturer Narrative
This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms#: 617147.No lot number was provided; therefore, device history record review could not be completed.No product sample was received; therefore, visual and functional testing could not be performed.Although no product was returned, an investigation was able to be conducted using the source files for the instruction of use(ifu).The alleged discrepancy was identified, and additional actions were submitted to the graphics labeling supervisor and graphics designer of the manufacturing site.The root cause was unable to be determined.Actions were taken to mitigate the reported issue: the manager, global technical communications was notified of the issue and a technical writer has been assigned to manage the translation updates for the product ifu.As a preventative measure, the translation vendor will be contracted to review all the languages to confirm that there are no additional errors.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA TRACHEOSTOMY TUBE, ADULT
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15598944
Report Number3012307300-2022-23875
Device Sequence Number2200699
Product Code JOH
UDI-Device Identifier15021312013825
UDI-Public15021312013825
Combination Product (Y/N)N
Initial Reporter CountryFI
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number67HA90
Device Catalogue Number67HA90
Device Lot NumberUNAVAILABLE
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/26/2022
Initial Report FDA Received Date10/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-