| Catalog Number |
46535RIM |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The suspect device is expected to return for investigation.A follow-up report will be submitted once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
The account alleges that during a percutaneous transluminal angioplasty of a patients superficial femoral artery [sfa] a catheter tip detached.The physician had acquired retrograde, arterial access and had negotiated the patient's ilio-femoral conduit under fluoroscopy with a guidewire and the 4f catheter.During over-the-wire catheter manipulations, the catheter tip detached but remained on the guidewire.The procedure was delayed.A vascular snare device was used to pin the detached tip to the wire during removal.The angioplasty procedure was completed post foreign body removal.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device was returned for evaluation.The complaint was confirmed, and the root cause is attributed to clinical use and forces imposed during use.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.
|
| |
|