• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER ORTHOPAEDIC MFG. LTD. ARTICULAR SURFACE MEDIAL CONGRUENT (MC) LEFT 12 MM HEIGHT; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER ORTHOPAEDIC MFG. LTD. ARTICULAR SURFACE MEDIAL CONGRUENT (MC) LEFT 12 MM HEIGHT; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Codes Difficult to Insert (1316); Peeled/Delaminated (1454)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/20/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Report source: japan.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported the patient underwent a total knee arthroplasty.During the procedure, the surgeon had difficulty seating the tibial insert in the tibial tray.After seating it in the tibial tray, the surgeon removed the insert from the tibial tray and noted the backside of the tibial insert peeled off.The surgeon attempted to seat the tibial insert again but the same issue occurred.The surgery was completed with a different tibial insert.No adverse events have been reported as a result of the malfunction.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Reported event was confirmed by review of photograph.Visual examination of the provided pictures shows only white poly string which looks like separated from the articular surface.Device history record was reviewed and no discrepancies relevant to the reported event were found.Review of complaint history found no additional related issues for this item and the reported part and lot combination.The root cause of the reported issue is attributed to user error because surgeon was hitting the articular surface during the insertion.A summary of the investigation was sent to the complainant conveying proper surgical technique and use of the device.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE MEDIAL CONGRUENT (MC) LEFT 12 MM HEIGHT
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15599176
Report Number3007963827-2022-00258
Device Sequence Number2215528
Product Code MBH
UDI-Device Identifier00889024469310
UDI-Public(01)00889024469310(17)261026(10)65197609
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K150090
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42512100812
Device Lot Number65197609
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/20/2022
Supplement Date Received by Manufacturer02/13/2023
Initial Report FDA Received Date10/13/2022
Supplement Report FDA Received Date02/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/11/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-