| Catalog Number |
397003-001 |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/20/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The companion hospital cart was not supporting a patient at the time of the event.The customer, a syncardia certified hospital, reported that the display on the companion hospital cart was "frozen", i.E., not recognizing touch.The hospital staff could not recalibrate the screen.
|
| |
|
Additional Manufacturer Narrative
|
|
The companion hospital cart will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.
|
| |
|
Additional Manufacturer Narrative
|
|
Device history record (dhr) review confirmed that companion hospital cart s/n 50062 was serviced and passed all functional testing prior to being released to finished goods.Visual inspection of external components found no abnormalities.Hospital card failed functional testing for acceptance at incoming inspection for sections all related to inability to interact with the lcd screen, confirming reported complaint.Additional testing included replacing the lcd module due to inability to interact with the lcd screen.Hospital cart passed all areas of functional testing with replacement module, including ground continuity testing.Failure investigation for this complaint confirmed he reported issue at incoming inspection.The customer complaint was replicated during tersting; the root cause of the reported inability to interact with the lcd screen was determined to be a faulty lcd module.Failure investigation identified no other failures or damage that could have contributed to the reported complaint.The hospital cart was not supporting a patient at time of complaint.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its evaluation and is closing this file.If new or additional information is received in the future, syncardia will file a follow-up mdr.
|
| |
|
Event or Problem Description
|
|
The companion hospital cart was not supporting a patient at the time of the event.The customer, a syncardia certified hospital, reported that the display on the companion hospital cart was "frozen", i.E., not recognizing touch.The hospital staff could not recalibrate the screen.
|
| |
|