BOSTON SCIENTIFIC CORPORATION ACQUIRE PULMONARY; BRONCHOSCOPE (FLEXIBLE OR RIGID) Back to Search Results
Model Number M00552350
Medical Device Problem Codes Leak/Splash (1354); Material Too Rigid or Stiff (1544); Activation, Positioning or Separation Problem (2906); Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/13/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Pma/510k: premarket / 510(k) #: k163248, k151895.(b)(4).
 
Event or Problem Description
It was reported to boston scientific corporation that an acquire pulmonary needle was used during an endobronchial ultrasound (ebus) procedure performed on (b)(6) 2022.During the procedure, the needle was inserted into the bronchial tree and the sheath was blocked and rigid.It had been slightly forced but it leaked, and it was no longer possible to adjust.After the procedure, it was noted the distal part of the bronchoscope was damaged.The procedure was not completed successfully due to the inability to adjust the sheath.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACQUIRE PULMONARY
Common Device Name
BRONCHOSCOPE (FLEXIBLE OR RIGID)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15601297
Report Number3005099803-2022-05779
Device Sequence Number2263224
Product Code EOQ
UDI-Device Identifier08714729986225
UDI-Public08714729986225
Combination Product (Y/N)N
Initial Reporter CountryIT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/21/2025
Device Model NumberM00552350
Device Catalogue Number5235
Device Lot Number0029635503
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/20/2022
Initial Report FDA Received Date10/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-