• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD ALARIS SMARTSITE GRAVITY SET; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD ALARIS SMARTSITE GRAVITY SET; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Model Number 10053857
Medical Device Problem Code No Flow (2991)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/20/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that an unspecified number of bd alaris smartsite gravity sets were clogged.The following information was provided by the initial reporter: staff had to go through a few iv gravity sets before they found one that worked.The tubing that connects will not flow once iv fluids are punctured and the dip chamber will not fill.
 
Additional Manufacturer Narrative
H.6.Investigation summary: no product or photo was returned by the customer.The customer complaint that it will not flow once iv fluids are punctured could not be verified due to the product not being returned for failure investigation.A device history record review for model 10053857 lot number 22049399 was performed.The search showed that a total of (b)(4) units in 1 lot number was built on 05may2022.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.Due to no sample being received, an investigation could not be performed and a root cause could not be determined.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.H3 other text : see h10.
 
Event or Problem Description
It was reported that an unspecified number of bd alaris smartsite gravity sets were clogged.The following information was provided by the initial reporter: staff had to go through a few iv gravity sets before they found one that worked.The tubing that connects will not flow once iv fluids are punctured and the dip chamber will not fill.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ALARIS SMARTSITE GRAVITY SET
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15601378
Report Number9616066-2022-01545
Device Sequence Number2200782
Product Code FPA
UDI-Device Identifier50885403234123
UDI-Public50885403234123
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K820278
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date04/29/2025
Device Model Number10053857
Device Catalogue Number10053857
Device Lot Number22049399
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/20/2022
Supplement Date Received by Manufacturer12/02/2022
Initial Report FDA Received Date10/13/2022
Supplement Report FDA Received Date12/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/29/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-