| Model Number |
24674 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/30/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The target lesion was located in a vein.A 10.0 x 40, 135cm mustang balloon catheter was advanced for dilatation.However, during inflation under normal working pressure, the balloon ruptured.The procedure was completed with another of the same device.There were no patient complications reported and the patient status was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: a mustang device, 14 atm was received for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.The rated burst pressure for this device is 14 atmospheres as per mustang specification.The returned device was attached to an encore inflation unit.Positive pressure was applied when liquid was observed to be leaking from a balloon pinhole located approximately 12mm proximal from the proximal markerband.An examination of the balloon material and markerbands identified no issues which could potentially have contributed to this complaint.A visual examination of the markerbands identified no issues which may potentially have contributed to the complaint incident.A visual and tactile examination identified no kinks or damage to the shaft which may have potentially contributed to the complaint incident.A visual and tactile examination identified no damage to the tip which may have potentially contributed to the complaint incident.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The target lesion was located in a vein.A 10.0 x 40, 135cm mustang balloon catheter was advanced for dilatation.However, during inflation under normal working pressure, the balloon ruptured.The procedure was completed with another of the same device.There were no patient complications reported and the patient status was stable.It was further reported that the 70% stenosed target lesion was located in a non-tortuous and non-calcified vein.The balloon was inflated once at 12 atmospheres for 3 seconds before it ruptured.
|
| |
|
Additional Manufacturer Narrative
|
|
B5 - updated describe event or problem.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The target lesion was located in a vein.A 10.0 x 40, 135cm mustang balloon catheter was advanced for dilatation.However, during inflation under normal working pressure, the balloon ruptured.The procedure was completed with another of the same device.There were no patient complications reported and the patient status was stable.It was further reported that the 70% stenosed target lesion was located in a non-tortuous and non-calcified vein.The balloon was inflated once at 12 atmospheres for 3 seconds before it ruptured.
|
| |
|