| Model Number |
45-751019 |
| Medical Device Problem Code |
Premature Separation (4045)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/05/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was stated to be available for return to the manufacturer for evaluation but has not yet been received.The coil detachment cannot be confirmed at this time.If the device is received at a later date, an investigation will be performed and a supplemental mdr will be submitted.The instructions for use (ifu) identifies premature coil detachment as a potential complication associated with the use of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that a snare was used to remove a coil after the coil pre-maturely detached inside a catheter.This coil detached when pulling the coil back to reposition the coil during an embolization just distal of the confluence of the ovarian, renal and lumbar veins to prevent reflux.The coil was advanced through the catheter and was approximately 3/4 of the way out of the catheter into the vein when the physician attempted to reposition the coil by pulling it back into the catheter.As it was pulled back, the coil detached pre-maturely inside the catheter.The physician tried to slowly advance the catheter to reposition the coil.The coil came out of the catheter and a portion of the coil was visible in the renal vein.The physician was able to snare the coil and successfully remove the coil from the patient.There was no reported patient impact or harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation of the returned coil system found the proximal connector kinked, the strain relief damaged, and the implant separated from the pusher.No indication using a detachment controller was found on the pusher's heater coil.The investigation found the pusher's monofilament with a tensile break shape at the tip which is consistent with the device experiencing excessive force that exceeded the strength of the monofilament causing the implant to separate from the pusher.The physical evaluation of the device could not identify the conditions or circumstances that led to the damage, but the damage is consistent with the device experiencing forces over specification.
|
| |
|
Search Alerts/Recalls
|