MENTOR TEXAS MENTOR MEMORYGEL XTRA BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Model Number |
SMPX560 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
06/09/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.A manufacturing record evaluation is in progress.Once completed, a supplemental report will be submitted.Reason for device explant and/or reoperation: breast prosthesis rupture.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a 49-year-old caucasian female patient underwent a breast reconstruction revision procedure with a mentor memorygel xtra breast implant 560cc gel breast prosthesis that ruptured post implantation.Rupture of the patient¿s left breast prosthesis was diagnosed by a healthcare professional.As a result, the patient underwent unilateral explantation and replacement with a mentor memorygel xtra breast implant 560cc gel breast prosthesis on (b)(6) 2022.
|
| |
|
Additional Manufacturer Narrative
|
|
On 19-oct-2022, mentor completed the investigation on the suspect medical device.Device evaluation summary: according to the information received, a rupture occurred in the breast implant.The product was returned to mentor for evaluation.Mentor conducted a visual inspection.Visual analysis of the returned sample revealed that the breast implant was found to be ruptured and received in two parts.As the authorization form for examination was not received the product evaluation lab couldn¿t identify a conclusion due to the specific nature of this rupture.The evaluation was limited to non-destructive testing.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Although no conclusion could be reached on the cause of the reported event, the instructions for use contain the following caution: rupture can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.The following things may cause the implant to rupture: damage by surgical instruments; stressing the implant during implantation and weakening it; folding or wrinkling of the implant shell; excessive force to the chest (e.G.During closed capsulotomy); trauma; compression during mammographic imaging; and severe capsular contracture.Breast implants may also simply wear out over time.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|