• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ALTATRACK GUIDEWIRE; WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ALTATRACK GUIDEWIRE; WIRE, GUIDE, CATHETER Back to Search Results
Model Number 884838070974
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Vascular Dissection (3160)
Date of Event 09/20/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This case was reviewed and investigated according to the manufacturer¿s policy.Date of birth not available.Not applicable for this device.Not applicable for this device.Study name - learn registry.The altatrack guidewire was not returned to the manufacturer for evaluation, thus no returned product investigation was performed.Do not apply to this submission.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
 
Event or Problem Description
It was reported that during a peripheral therapeutic procedure, a altatrack berenstein catheter was loaded over an altatrack guidewire.During advancement, resistance was encountered while cannulating in the right renal artery.A dissection was noted, which was then covered with a stent.The physician stated that the guidewire caused the dissection, but there was no device malfunction.The patient was presumed in stable condition and discharged from the hospital.This adverse event is being submitted because the altatrack guidewire caused a dissection and required stent placement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTATRACK GUIDEWIRE
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 4-6
best 5684
NL  5684
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 4-6
best 5684
NL   5684
Manufacturer Contact
katarzyna feliksik
veenpluis 4-6
best 5684
NL   5684
MDR Report Key15601895
Report Number3003768277-2022-10501
Device Sequence Number16710730
Product Code DQX
UDI-Device Identifier00884838070974
UDI-Public(01)00884838070974(17)230719(10)6998296
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K201081
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Report Source Study,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/13/2022
Report Date (Section F) 10/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/19/2023
Device Model Number884838070974
Device Catalogue Number432207013481
Device Lot Number6998296
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date09/21/2022
Device Age2 MO
Event Location Hospital
Date Report to Manufacturer10/13/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date10/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/19/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MEDTRONIC: 7F TOURGUIDE INTRODUCER SHEATH; PHILIPS: ALTATRACK 5.5F BRN CATHETER
Outcome Attributed to Adverse Event Required Intervention;
Patient Age80 YR
Patient SexFemale
Patient Weight79 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
Date Report Sent to FDA10/13/2022
-
-