Model Number 190610
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 09/01/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Clinical investigation: a temporal relationship exists between hd therapy with the fresenius 2008k2 hemodialysis machine and the patient¿s event.The machine will not be returned to manufacturer for device analysis.However, it was reported a biomedical engineer performed a safety check on the machine post incident without any reported allegations of a device malfunction or product issue.The machine was returned to use without any reported issues.Based on the very limited information, the exact cause for the patient incident can not be determined.However, it was reported by the biomedical engineer that the patient had several comorbid conditions that contributed to the event.The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Event or Problem Description
It was reported the fresenius 2008k2 hemodialysis machine had a safety check performed due to an unspecified patient incident.Additional follow-up was conducted with a biomedical engineer.However, very limited details about the patient event were available.It was stated an unknown patient coded while on the fresenius 2008k2 hemodialysis machine.It was reported as part of clinic policy, the machine is put through a safety check after any incident.The biomedical engineer reported the machine was not suspected to have caused any harm to the patient.It was stated the patient incident was due to patient comorbidities and not related to fresenius product.Per the biomedical engineer, the fresenius 2008k2 hemodialysis machine passed all tests and has returned to use without any issues.No further information was available.
 
Event or Problem Description
It was reported the fresenius 2008k2 hemodialysis machine had a safety check performed due to an unspecified patient incident.Additional follow-up was conducted with a biomedical engineer.However, very limited details about the patient event were available.It was stated an unknown patient coded while on the fresenius 2008k2 hemodialysis machine.It was reported as part of clinic policy, the machine is put through a safety check after any incident.The biomedical engineer reported the machine was not suspected to have caused any harm to the patient.It was stated the patient incident was due to patient comorbidities and not related to fresenius product.Per the biomedical engineer, the fresenius 2008k2 hemodialysis machine passed all tests and has returned to use without any issues.No further information was available.
 
Additional Manufacturer Narrative
Plant investigation: no parts were returned to the manufacturer for physical evaluation.However, an on-site evaluation was performed by a fresenius field service technician (fst).The ultrafiltration (uf) checklist was performed including calibration of the uf pump.It was documented that the machine passed all functional testing.A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The investigation was not able to confirm the reported failure mode.A causal relationship between the objective evidence and the alleged event could not be determined, therefore the alleged event is unconfirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2008K2 HEMODIALYSIS SYS. OLC/DIASAFE PLS
Common Device Name
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
6174175172
MDR Report Key15602326
Report Number0002937457-2022-01720
Device Sequence Number16713866
Product Code KDI
UDI-Device Identifier00840861100859
UDI-Public00840861100859
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K994267
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number190610
Device Catalogue Number190610
Was Device Available for Evaluation? Yes
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/27/2022
Supplement Date Received by Manufacturer10/21/2022
Initial Report FDA Received Date10/13/2022
Supplement Report FDA Received Date11/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/04/2009
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-