• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON BD MICROTAINER® CONTACT-ACTIVATED LANCET; MANUAL SURGICAL INSTRUMENT FOR GENERAL USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON BD MICROTAINER® CONTACT-ACTIVATED LANCET; MANUAL SURGICAL INSTRUMENT FOR GENERAL USE Back to Search Results
Catalog Number 366592
Medical Device Problem Code Delivered as Unsterile Product (1421)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/06/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Oem manufacturer: the manufacturing location for this product is (b)(4).This site is an oem manufacturing site.(b)(4).
 
Event or Problem Description
It was reported when using the bd microtainer® contact-activated lancet that the cap was disassembled when it was removed.This event occurred two times.The following information was provided by the initial reporter.The customer stated: ¿the customer's verbatim report about two defective products is that the cap was disassembled when it was removed.¿.
 
Event or Problem Description
It was reported when using the bd microtainer® contact-activated lancet that the cap was disassembled when it was removed.This event occurred two times.The following information was provided by the initial reporter.The customer stated: ¿the customer's verbatim report about two defective products is that the cap was disassembled when it was removed.¿.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d9: device available for eval?: yes.D9: returned to manufacturer on: 2022-10-19.H.6.Investigation summary: bd received 2 samples and 4 photos for investigation.The photos were reviewed and the indicated failure mode for disassembled cap was observed.Additionally, the customer samples along with retention samples from bd inventory, were evaluated by visual examination and the issue of disassembled cap was observed.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint has been confirmed for the indicated failure mode disassembled cap.Bd determined that the root cause of the indicated failure mode was the tab cap was not assembled during the assembly process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MICROTAINER® CONTACT-ACTIVATED LANCET
Common Device Name
MANUAL SURGICAL INSTRUMENT FOR GENERAL USE
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15602378
Report Number2243072-2022-01762
Device Sequence Number16723314
Product Code FMK
UDI-Device Identifier50382903665926
UDI-Public50382903665926
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date09/30/2026
Device Catalogue Number366592
Device Lot NumberB9F69S7
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/06/2022
Supplement Date Received by Manufacturer11/14/2022
Initial Report FDA Received Date10/13/2022
Supplement Report FDA Received Date11/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/28/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-