| Model Number |
21587700 |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Ref sf: (b)(4).On (b)(6) 2022: it was reported hearing aid charger and usb-c connector melted and "fused" together so the usb-c cable could not be removed and cable broke.Only original equipment was used.The rechargable hearing aids with the charger were purchased on (b)(6) 2021.The date of the incident is not available.Device is out of warranty, the clinic purchased a new device.The date of the event is not availble.No patient involvment or harm to the user reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturers reference#: (b)(4).Based on the available information, the malfunction may lead to, or might led to, death or a serious injury if it were to recur or the event / incident did lead to, or might have led to, death or a serious injury and indicates need for immediate reporting.Technical investigation concluded: device history register review have been completed.No deviation or changes were found during manufacturing of the device.The clinical investigation concluded: based on these results from previous evaluations, clinical conclusion on this case is that there is a potential risk of heat dissipation based on the findings.From the previous investigations it is concluded that the high temperatures are limited to places difficult to touch, such as the back of the charger.Clinical conclusion on the case is that, even though unpleasant, it is evaluated unlikely that the device in failure mode would cause harm, if the event was to reoccur, because of the unlikely event of contacting the difficult to touch location.Risk register will be updated to include warning/caution to the users.Further investigation pending, to be submitted when completed.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturers reference #(b)(4).Based on the available information, the malfunction may lead to, or might led to, death or a serious injury if it were to recur or the event / incident did lead to, or might have led to, death or a serious injury and indicates need for immediate reporting.Technical investigation concluded: device history register review have been completed.No deviation or changes were found during manufacturing of the device.The clinical investigation concluded: based on these results from previous evaluations, clinical conclusion on this case is that there is a potential risk of heat dissipation based on the findings.From the previous investigations it is concluded that the high temperatures are limited to places difficult to touch, such as the back of the charger.Clinical conclusion on the case is that, even though unpleasant, it is evaluated unlikely that the device in failure mode would cause harm, if the event was to reoccur, because of the unlikely event of contacting the difficult to touch location.Risk register will be updated to include warning/caution to the users.Further investigation pending, to be submitted when completed.
|
| |
|
Event or Problem Description
|
|
Ref (b)(4).On 13sep2022: it was reported hearing aid charger and usb-c connector melted and "fused" together so the usb-c cable could not be removed and cable broke.Only original equipment was used.The rechargable hearing aids witht he charger were purchased on (b)(6) 2021.The date of the incident is not available.Device is out of warranty, the clinic purchased a new device.The date of the event is not availble.No patient involvment or harm to the user reported.On 09nov2022: no new information recieved.No follow up is expected.
|
| |
|
Event or Problem Description
|
|
Ref sf (b)(4).(b)(6) 2022: it was reported hearing aid charger and usb-c connector melted and "fused" together so the usb-c cable could not be removed and cable broke.Only original equipment was used.The rechargable hearing aids with the charger were purchased on (b)(6) 21.The date of the incident is not available.Device is out of warranty, the clinic purchased a new device.The date of the event is not available.No patient involvement or harm to the user reported.09nov2022: no new information received.No follow up is expected.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturers reference # sf (b)(4).Based on the available information, the malfunction may lead to, or might led to, death or a serious injury if it were to recur or the event / incident did lead to, or might have led to, death or a serious injury and indicates need for immediate reporting.Technical investigation concluded: device history register review have been completed.No deviation or changes were found during manufacturing of the device.The clinical investigation concluded: based on these results from previous evaluations, clinical conclusion on this case is that there is a potential risk of heat dissipation based on the findings.From the previous investigations it is concluded that the high temperatures are limited to places difficult to touch, such as the back of the charger.Clinical conclusion on the case is that, even though unpleasant, it is evaluated unlikely that the device in failure mode would cause harm, if the event was to reoccur, because of the unlikely event of contacting the difficult to touch location.Risk register will be updated to include warning/caution to the users.Further investigation pending, to be submitted when completed.09nov2022: additional information/corrections: b3 - best estimate for data of event added, b5 - comment added, d8 - corrected, g6 - corrected, h2 - corrected, h8 - corrected.Device investigation pending, to be submitted when available.09dec2022.Additional information/corrections: b4 - date of this report updated, d9 - updated, e2 and e3 - initial reporter updated, g2 - report source updated, h3 - type of investigation code added, investigation findings and investigation conclusion codes updated.Manufacturer's investigation conclusion: trended case device investigation concluded: the usb-c connector has broken down mechanical.It is a known problem with usb connectors, that in rare cases wear or dirt/moist can create a low ohmic path from vbus to gnd, that potential can cause a heating inside of the connector.The clinical investigation concluded (updated): based on these results from previous evaluations, clinical conclusion on this case is that there is a potential risk of heat dissipation based on the findings.From the previous investigations it is concluded that the high temperatures are limited to places difficult to touch, such as the back of the charger.Clinical conclusion on the case is that, even though unpleasant, it is evaluated unlikely that the device in failure mode would cause harm, if the event was to reoccur, because of the unlikely event of contacting the difficult to touch location.Risk register has been updated to include warning/caution to the users.Manufacture's investigation is final.Consider this a final report.
|
| |
|