• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GN HEARING A/S GN RESOUND; HEARING AID, AIR CONDUCTION WITH WIRELESS TECHNOLOGY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GN HEARING A/S GN RESOUND; HEARING AID, AIR CONDUCTION WITH WIRELESS TECHNOLOGY Back to Search Results
Model Number 21587700
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
Ref sf: (b)(4).On (b)(6) 2022: it was reported hearing aid charger and usb-c connector melted and "fused" together so the usb-c cable could not be removed and cable broke.Only original equipment was used.The rechargable hearing aids with the charger were purchased on (b)(6) 2021.The date of the incident is not available.Device is out of warranty, the clinic purchased a new device.The date of the event is not availble.No patient involvment or harm to the user reported.
 
Additional Manufacturer Narrative
Manufacturers reference#: (b)(4).Based on the available information, the malfunction may lead to, or might led to, death or a serious injury if it were to recur or the event / incident did lead to, or might have led to, death or a serious injury and indicates need for immediate reporting.Technical investigation concluded: device history register review have been completed.No deviation or changes were found during manufacturing of the device.The clinical investigation concluded: based on these results from previous evaluations, clinical conclusion on this case is that there is a potential risk of heat dissipation based on the findings.From the previous investigations it is concluded that the high temperatures are limited to places difficult to touch, such as the back of the charger.Clinical conclusion on the case is that, even though unpleasant, it is evaluated unlikely that the device in failure mode would cause harm, if the event was to reoccur, because of the unlikely event of contacting the difficult to touch location.Risk register will be updated to include warning/caution to the users.Further investigation pending, to be submitted when completed.
 
Additional Manufacturer Narrative
Manufacturers reference #(b)(4).Based on the available information, the malfunction may lead to, or might led to, death or a serious injury if it were to recur or the event / incident did lead to, or might have led to, death or a serious injury and indicates need for immediate reporting.Technical investigation concluded: device history register review have been completed.No deviation or changes were found during manufacturing of the device.The clinical investigation concluded: based on these results from previous evaluations, clinical conclusion on this case is that there is a potential risk of heat dissipation based on the findings.From the previous investigations it is concluded that the high temperatures are limited to places difficult to touch, such as the back of the charger.Clinical conclusion on the case is that, even though unpleasant, it is evaluated unlikely that the device in failure mode would cause harm, if the event was to reoccur, because of the unlikely event of contacting the difficult to touch location.Risk register will be updated to include warning/caution to the users.Further investigation pending, to be submitted when completed.
 
Event or Problem Description
Ref (b)(4).On 13sep2022: it was reported hearing aid charger and usb-c connector melted and "fused" together so the usb-c cable could not be removed and cable broke.Only original equipment was used.The rechargable hearing aids witht he charger were purchased on (b)(6) 2021.The date of the incident is not available.Device is out of warranty, the clinic purchased a new device.The date of the event is not availble.No patient involvment or harm to the user reported.On 09nov2022: no new information recieved.No follow up is expected.
 
Event or Problem Description
Ref sf (b)(4).(b)(6) 2022: it was reported hearing aid charger and usb-c connector melted and "fused" together so the usb-c cable could not be removed and cable broke.Only original equipment was used.The rechargable hearing aids with the charger were purchased on (b)(6) 21.The date of the incident is not available.Device is out of warranty, the clinic purchased a new device.The date of the event is not available.No patient involvement or harm to the user reported.09nov2022: no new information received.No follow up is expected.
 
Additional Manufacturer Narrative
Manufacturers reference # sf (b)(4).Based on the available information, the malfunction may lead to, or might led to, death or a serious injury if it were to recur or the event / incident did lead to, or might have led to, death or a serious injury and indicates need for immediate reporting.Technical investigation concluded: device history register review have been completed.No deviation or changes were found during manufacturing of the device.The clinical investigation concluded: based on these results from previous evaluations, clinical conclusion on this case is that there is a potential risk of heat dissipation based on the findings.From the previous investigations it is concluded that the high temperatures are limited to places difficult to touch, such as the back of the charger.Clinical conclusion on the case is that, even though unpleasant, it is evaluated unlikely that the device in failure mode would cause harm, if the event was to reoccur, because of the unlikely event of contacting the difficult to touch location.Risk register will be updated to include warning/caution to the users.Further investigation pending, to be submitted when completed.09nov2022: additional information/corrections: b3 - best estimate for data of event added, b5 - comment added, d8 - corrected, g6 - corrected, h2 - corrected, h8 - corrected.Device investigation pending, to be submitted when available.09dec2022.Additional information/corrections: b4 - date of this report updated, d9 - updated, e2 and e3 - initial reporter updated, g2 - report source updated, h3 - type of investigation code added, investigation findings and investigation conclusion codes updated.Manufacturer's investigation conclusion: trended case device investigation concluded: the usb-c connector has broken down mechanical.It is a known problem with usb connectors, that in rare cases wear or dirt/moist can create a low ohmic path from vbus to gnd, that potential can cause a heating inside of the connector.The clinical investigation concluded (updated): based on these results from previous evaluations, clinical conclusion on this case is that there is a potential risk of heat dissipation based on the findings.From the previous investigations it is concluded that the high temperatures are limited to places difficult to touch, such as the back of the charger.Clinical conclusion on the case is that, even though unpleasant, it is evaluated unlikely that the device in failure mode would cause harm, if the event was to reoccur, because of the unlikely event of contacting the difficult to touch location.Risk register has been updated to include warning/caution to the users.Manufacture's investigation is final.Consider this a final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GN RESOUND
Common Device Name
HEARING AID, AIR CONDUCTION WITH WIRELESS TECHNOLOGY
Manufacturer (Section D)
GN HEARING A/S
lautrupbjerg 7
ballerup, 2750
DA  2750
Manufacturer Contact
karen frisk
lautrupbjerg 7
ballerup, 2750
DA   2750
MDR Report Key15604245
Report Number3005650109-2022-00049
Device Sequence Number2278086
Product Code OSM
UDI-Device Identifier05708296200337
UDI-Public05708296200337
Combination Product (Y/N)N
Initial Reporter StateAK
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number21587700
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/13/2022
Supplement Date Received by Manufacturer09/13/2022
09/13/2022
Initial Report FDA Received Date10/14/2022
Supplement Report FDA Received Date11/11/2022
12/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/01/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-